An Evaluation of Fiberglass Cast Application Techniques
Cylindrical casts constructed from 2, 3, 4, and 5 inch widths of fiberglass casting material were evaluated for strength in bending using an unstable fracture model. Five-inch casting material produced the strongest casts.
The effect of longitudinal splinting with 2-inch fiberglass tape was evaluated in cylindrical casts constructed from 4-inch tape. Casts with splints on the compression side or with splints on both the tension and compression sides were stronger in bending than casts with a single splint on the tension side or casts without splints.